DVAP-3

Dewar-Style Ambient Air Vaporizer

Largest · Most Flow

Highest-capacity DVAP model for demanding intermittent Dewar and liquid-cylinder applications.

Flow Ratings

SCFH @ 300 PSI, LN2 / Air @ 70F, sustained flow by run time

Flow ratings for the DVAP-3 vaporizer
Run TimeSustained Flow
4 hour3,600 SCFH
8 hour2,800 SCFH
24 hour2,240 SCFH

Specifications

Specifications for the DVAP-3 vaporizer
ModelDVAP-3
MAWP600 PSIG
Service TypeIntermittent (Dewar / single-cylinder duty)
Standard FittingsStandard easy-connect fittings (typical)
Connection SpecOn dimensional drawing
Optional ConnectionsAvailable on request
MaterialAluminum
Oxygen ServiceCleaned, sealed & tagged
ComplianceASME B31.3
Unit Dimensions (H x W x D)65.7 x 26 x 19 in
Dry Weight (approx.)134 lb
Packaged Dimensions & Weight70 x 24 x 33 in · 240 lb
AvailabilityCurrent availability and release timing are confirmed on every order or quotation.

Flow basis. Flow ratings are sustained nitrogen / air ratings measured at 300 PSI with LN2 / air at 70F, listed by run time (4, 8, and 24 hour). Actual performance varies with ambient temperature, humidity, gas service, and duty cycle. Consult FWM for application-specific sizing before purchase.

Service. DVAP units are sized for intermittent Dewar and single-cylinder withdrawal duty. For continuous high-flow service, the FWM AV Series ambient vaporizers are typically the better fit.

Connections. Standard easy-connect 90-degree fittings are supplied. Optional connection sizes and types are available on request; note your required connection on the quote and FWM will confirm it during engineering review.

Dimensions. Unit dimensions (H x W x D) are measured from the FWM 3D model; dry weights are approximate. Final dimensions are confirmed at engineering review, and a dimensioned reference drawing is available on request.

Compatibility. DVAP units are compatible with many standard cryogenic liquid cylinders and Dewars. Confirm cylinder compatibility with FWM for your specific gas service and vessel.

Oxygen service. All DVAP units are cleaned, sealed, and tagged for oxygen service. Final oxygen-service cleaning and documentation are confirmed with FWM for the delivered configuration.

Documentation. Datasheets, drawings, and documentation packages are available on request and confirmed during engineering review.

Information FWM Needs to Confirm Selection

  • Gas service and required flow rate
  • Required run time (4 / 8 / 24 hour)
  • Liquid cylinder or Dewar type
  • Inlet and outlet pressure
  • Connection size and type
  • Site ambient design temperature
  • Oxygen-service cleaning and documentation needs
  • In-stock vs. scheduled delivery
Start an Order for DVAP-3(603) 578-3366info@fwminc.com